by Lawrence W. Bassett MD
A broad range of image-guided biopsy techniques are now in clinical use, offering alternatives to the more costly and disfiguring open surgical biopsy for properly elected patients.
Breast-biopsy options have changed considerably over time. In
part, biopsy methods have changed in order to keep up with new
developments in the detection and diagnosis of breast cancer.
Advances in imaging have been prominent among these developments,
and the broadest range of biopsy techniques yet seen in medicine is
now in clinical use. While open surgical biopsies are still
commonly performed, image-guided needle biopsies are preferable
under many circumstances.
PALPABLE MASSES
At one time, breast lumps found during physical examinations or
noted during a patient's self-examination were usually removed
completely during excisional biopsy. This procedure was performed
in an operating room while the patient was under general
anesthesia, and palpation was used to guide the surgeon to the mass
to be excised. If, during surgery, it was possible to remove only a
portion of the area involved, the procedure was referred to as
incisional biopsy.
Open biopsy can be associated with significant cosmetic defects.
While these can be minimized by highly skilled, well trained
surgeons, a poor cosmetic result may still be inevitable if the
mass to be excised is large. The use of image-guided biopsy is the
subject of controversy where large lesions are concerned, but it
may still be preferred for palpable masses because its use can
ensure that the mass has been sampled properly, with the resulting
histological information being helpful in later patient care. For
example, most surgeons want to be sure a cancer is invasive before
doing a biopsy of the axillary nodes.
If the patient who has a palpable benign mass on imaging has
decided before biopsy that she wants to have the mass removed, then
surgical excision remains the procedure of choice and a core biopsy
is not indicated. Any other biopsy method would simply add an
unnecessary expense to the patient's management. Pressures favoring
cost-effectiveness in medicine continue to mount, so it is
important to avoid subjecting the patient to two procedures where
one will do.
MAMMOGRAPHY FINDINGS
With the advent of mammography screening, the surgeon's task in
performing an open breast biopsy became much more difficult.
Palpable masses had once been the only lesions subjected to
excisional biopsy, but far smaller abnormalities could be detected
mammographically. For this reason, the surgeon could no longer
count on feeling the suspicious area; this made it hard to be
certain that the correct area had been sampled.
In response to this problem, techniques were developed to help
the surgeon localize the suspicious lesion more reliably. An
intricate system of mapping was first used to show the relationship
between the abnormal area and the surface landmarks of the breast.
A second technique, specimen radiography, involved radiographic
examination of the biopsy specimen to determine whether it matched
the abnormal mammography findings; this improved the odds that the
intended target had actually been removed for examination.
Although these two techniques were viewed as helpful, it was
still necessary to excise a large amount of tissue in order to be
certain that the lesion had been removed. At the same time, the use
of mammography spread. With smaller abnormalities being detected,
the number of excisional biopsies naturally increased. In order to
avoid the removal of excessive amounts of breast tissue,
radiologists focused on developing better ways of pinpointing
lesions for surgical removal.
Toward this end, guided by mammograms, radiologists began
placing radiopaque markers directly over suspicious areas. These
markers on the breast's surface were then used during surgery to
ensure better localization of lesions. Soon after, the injection of
dye into the lesion and the needle track leading to it were
employed to enhance surgical guidance. To increase precision still
further, dye-injection needles were left in place until surgery.
This was followed by the development of hookwire systems that could
hold needles or wire in place very near the area of questionable
findings. This permitted the amount of excised tissue to be
decreased while the accuracy of sampling was retained.
All of these image-guided methods still required open biopsy,
which is the gold standard against which later developments would
be measured. When a suspicious mass is removed in its entirety (as
verified using specimen radiography), sampling error is almost
nonexistent. With the involvement of two experts-the experienced
radiologist who localizes a lesion and the experienced surgeon who
removes it-the failure rate for the excision of nonpalpable
abnormalities is less than 2% (although the medical literature
varies on this point).
FINE-NEEDLE ASPIRATION
The first alternative to open biopsy, called fine-needle
aspiration cytology (FNAC) or fine-needle aspiration biopsy (FNAB),
was developed in Europe for use with palpable masses. Its logical
application to nonpalpable lesions under stereotactic guidance was
introduced in Sweden in 1979 by Nordenström and Zajieck.
During the decade that followed, the technique was widely used in
Europe to evaluate most abnormal mammographic findings, and it
replaced open biopsy to a large degree. Ultrasound was used to
guide FNAC during the latter half of the 1980s. The fine-needle
aspiration method, however, met with only limited success in the
United States. Several factors contributed to this delay in
adoption. Too few skilled cytologists were available to evaluate
fine-needle aspiration specimens. In addition, needle placement was
not always precise. There was also some hesitation to use the
fine-needle because of possible exposure to medical liability.
Furthermore, because of reports of relatively large numbers of
false negatives, specimens obtained through FNAC did not permit
interpreters to determine whether the abnormalities noted
represented carcinoma in situ, invasive carcinoma, or atypical
hyperplasia. The diagnostic accuracy reported for the procedure in
the literature varied widely, as well.
In an attempt to establish a uniform approach to FNAB of the
breast, the National Cancer Institute of the National Institutes of
Health convened a consensus conference in September 1996.
Guidelines for performing FNAB and interpreting its results were
established by the conference. The group recommended the use of two
to four passes of up to 15 up-and-down motions each, employing a
22-gauge to 25-gauge needle.
STEREOTACTIC EQUIPMENT
Accuracy, in needle biopsy, clearly depends on imaging for
needle placement. Having mammographic guidance available from two
different angles greatly improves the accuracy of needle placement
by identifying the depth of the lesion within the breast. This
stereotactic biopsy method is supported by two primary equipment
configurations, upright and prone. In prone stereotactic biopsy,
the patient lies, face downward, on a table dedicated to this
purpose. The breast, accessible through an opening in the table, is
compressed and held while the biopsy is performed by the
radiologist and technologist, who are typically seated beneath the
elevated table. Use of the prone table prevents vasovagal reactions
such as loss of consciousness and can permit examinations to be
completed more rapidly. This type of system is now more widely used
but its more expensive than an upright unit. The prone table also
requires more space.
Upright stereotactic biopsy equipment that fits on a mammography
unit is typically less expensive, and it often allows use of the
space for mammography examinations between biopsy patients. The
radiologist and technologist have less room to work, however, and
vasovagal reactions are possible. Some women may also find it
difficult to remain still during the procedure if they are in a
seated position.
CORE-NEEDLE BIOPSY
Core-needle biopsy (CNB) using an automated large-core (ALC) gun
was introduced in Sweden by Lindgren in 1982. It was adapted to
stereotactic breast biopsies in 1990, by Parker. Later, ultrasound
guidance for breast core-needle biopsy was introduced. The two
guidance methods led to a revolution in US breast-biopsy methods,
with CNB reaching wide dissemination and rapid adoption.
Breast CNB uses 18-gauge, 16-gauge, or the preferred 14-gauge
needles that have a throw of 1 cm to 2.5 cm into the breast in
order to obtain specimens. Typically, five or fewer specimens are
obtained for masses and 10 or more are needed to evaluate
calcifications. CNB provides larger samples than FNAB, and permit
pathologists to determine whether a carcinoma is in situ or
invasive. A few CNB histological diagnoses, however, require that
excisional biopsy be performed
VACUUM-ASSISTED DEVICES
A disposable probe and a reusable probe driver are combined to
form the vacuum-assisted device (VAD) used to obtain specimens of
breast tissue. A cutting unit within the probe captures tissue that
is then removed from the breast using vacuum aspiration. The probe
can be guided stereotactically (as originally designed) or using
ultrasound. Once the target specimen has been retrieved, a
microclip can be inserted to mark the biopsy site.
Vacuum-assisted devices use 14-gauge to 11-gauge needles; the
larger needles produce a 70% increase in specimen diameter that is
especially helpful in the assessment of calcifications. Specimen
retrieval is faster, and samples are more contiguous. The probes
can be expensive, however, and there may be additional costs
involved if it is necessary to upgrade an older stereotactic system
to handle vacuum-assisted biopsies.
MICROCLIP PLACEMENT and DOCUMENTATION
At the end of a stereotactic vacuum-assisted biopsy, a 2-mm
stainless steel clip can be placed in the suspicious area directly
through the 11-gauge probe. A smaller clip or coil can be
introduced through the 14-gauge needle. If the lesion has been
entirely removed during the biopsy, the clip or coil can then serve
to identify the biopsy site, as well as to identify the tumor bed
if nonsurgical cancer therapy later proves necessary.
Biopsy documentation through imaging is important not only for
medicolegal reasons, but to ensure that the pathology and imaging
results refer to the same lesion. During the biopsy, imaging
documentation is used to verify that the specimens came from the
target area.
In our practice, for stereotactic biopsy, documentation calls
for targeting images at 0° and at ±15° (stereo
pairs), along with stereo images before and after firing. Another
image is obtained to show immediate changes following the biopsy.
Specimen radiography then confirms that the specimen contains
calcifications, and postbiopsy mammograms show the location of the
microclip, if one was inserted.
For documentation of ultrasound-guided biopsy, prefire images
showing the needle proximal to the lesion should be followed by
postfire images that show the needle traversing the lesion
area.
CONCLUSION
Whether a radiologist uses ultrasound or stereotactic
mammography to guide needle biopsy is often a matter of preference.
Ultrasound may be preferred for preoperative needle localization or
for needle biopsy of palpable masses. It is more rapid, it permits
real-time imaging, and it is likely to be more comfortable for the
patient. For nonpalpable lesions visible only mammographically,
stereotactic guidance is necessary.
Needle biopsy techniques obviously procure specimens at a lower
cost and are less invasive than open biopsy methods. The biopsy
itself may be easier and less expensive, but patients who would not
have been sent for open biopsies should not be sent for needle
biopsies, either. The recommendation for biopsy threshold should
not be lowered simply because we have new biopsy devices, without
evidence that patient care and outcomes would be improved. No
savings in cost or decrease in patient worry can be produced if
biopsy is simply recommended more often. Better care at a lower
cost, however, will be obtained when needle biopsy replaces open
biopsy for properly selected patients.
Additional Reading
• Azevado E, Svane G, Aver G. Stereotactic fine needle
biopsy in 2594 mammographically detected nonpalpable lesions.
Lancet. 1989;1:1033-1036.
• Bassett LW, Winchester DP, Caplan RB, et al.
Stereotactic breast CNB: report of the Joint Task Force of ACR,
ACS, COAP. CA Cancer J Clin. 1997;47:171-190.
• Brenner RJ, Fajardo L, Fisher PR, et al. Percutaneous
CNB: effect of operator experience and number of samples on
accuracy. AJR Am J Roentgenol. 1996;166:341-346.
• Burbank F, Forcier N. Tissue marking clip for
stereotactic breast biopsy: initial placement accuracy, long-term
stability, and usefulness as a guide for wire localization.
Radiology. 1997;205:407-415.
• Dershaw DD, Morris EA, Liberman L, Abramson AF.
Nondiagnostic stereo CNB: results of rebiopsy. Radiology. 1996;
198:323-325.
• Liberman L, Dershaw DD, Rosen PP, et al. Stereotaxic
14-gauge breast biopsy: how many core biopsy specimens are needed?
Radiology. 1994;192:793-795
• Liberman L, Evans WP, Dershaw DD, et al. Radiography of
calcifications in stereo CNB specimens. Radiology.
1994;190:223-225.
• Löfgren M, Andersson I, Lindholm K. Stereotactic
fine-needle aspiration for cytologic diagnosis of nonpalpable
breast lesions. AJR Am J Roentgenol. 1990;154:1191-1195.
• Morrow M. Indications for the use of stereotactic core
biopsy of the breast. Curr Probl Gen Surg. 1996;13:1-8.
• Parker SH, Lovin JD, Jobe WE, et al. Stereotactic breast
biopsy with a biopsy gun. Radiology. 1990;176:741-747.
• Parker SH, Klaus AJ. Performing CNB with a directional,
vacuum-assisted biopsy instrument. Radiographics.
1997;17:1233-1252.
• Philpotts LE, Shaheen NA, Carter D, Lange RC, Lee CH.
Comparison of rebiopsy rates after stereotactic core needle biopsy
of the breast with 11-gauge vacuum suction probe versus 14-gauge
needle and automatic gun. AJR Am J Roentgenol.
1999;172:683-687.
• Uniform approach to breast fine-needle aspiration
biopsy. Recommendations of the National Cancer Institute sponsored
conference, Bethesda, Maryland, September 9-10, 1996. Diagn
Cytopathol. 1997;16:295-311.
Lawrence W. Bassett, MD, is Iris Cantor Professor of Breast Imaging, UCLA School of Medicine, Los Angeles. This article is based on his presentation at the Fifth Postgraduate Course of the Society of Breast Imaging on May 16, 2001, in San Diego.