by Charles E. Willis PhD
In addressing what it takes to produce high-quality computed radiography images, the author dispels commonly held CR fallacies
|
With the passage of two decades since its introduction, computed
radiography (CR) has become a mainstream technology for acquiring
ordinary radiographic projections in digital form. CR is extremely
flexible, promising to replace virtually all conventional
screen-film radiography, including mammography, scoliosis, and
panorex studies. CR is especially suited to bedside radiographic
examinations, traditionally the worst images in the hospital
because of the exigent circumstances under which the examinations
are performed. CR's wide latitude and automatic density adjustment
dramatically improve the consistency of bedside radiographs. The
appearance of the CR image can be modified as desired after
acquisition. The CR image can be distributed virtually anywhere
electronically, viewed simultaneously by multiple care providers,
and reprinted, as necessary. The CR image is acquired on reusable
image media (RIM), which can be erased and imaged thousands of
times, eliminating the need for film and chemicals.
QC STILL CRITICAL
The success of CR leads to the misconception that quality
control (QC) processes are no longer necessary. Early adopters of
CR claimed that retake rates decreased to zero. In actual fact, QC
processes for CR are no less important than they are for
conventional screen-film radiography, and must be modified to take
into consideration the unique characteristics of CR technology.
Fallacy No.1. No more bad images with CR. Traditionally, the
primary cause of rejected radiographic images is mispositioning.
Automation has not been invented to correct for mispositioning,
patient motion, inadequate inspiration, wrong examination
performed, wrong patient examined, improper collimation, double
exposure, or the many potential artifacts that can interfere with
the proper acquisition of a radiographic projection. It would be
naive to believe that all these classic causes of rejected images
would suddenly cease with CR. Indeed, analysis demonstrates that
rejected images occur about as frequently in CR institutions from
roughly the same causes as in conventional departments.
Unrealistically low reject rates are often the result of
inattention to the erasure of substandard electronic images (see
Figure 1).
Figure 1. Image resulting from double-exposure of cassette. This error occurs in conventional screen-film radiography and CR, illustrating Fallacy No. 1.
|
Fallacy No. 2. CR is "plug-and-play." QC begins with appropriate
acceptance testing of each CR system before it is introduced into
clinical service. During installation, it is important to verify
that all hardware, software, and durable goods that were purchased
have arrived in good condition. Every component needs to be
properly configured and integrated into the hospital's imaging
operation, and testing should be performed to assure this. Every CR
vendor has adjustment and calibration procedures specified in their
service manuals. At an absolute minimum, these calibrations should
be performed, verified, and documented. The manner in which patient
demographic information is associated with the CR image must be
configured and tested. The infinite variability with which the CR
image can be acquired and modified before transmission and display
allows a variety of errors in configuration by human operators (see
Figure 2).
TRAINING IS KEY
Knowledgeable, skilled technologists must perform QC processes.
Most technologist training programs do not address CR, and most
technologists have no outside experience with CR. This implies that
technologists must be trained about CR technology, initially by
vendor applications personnel. Additional training on local
practice, policies, and procedures must be developed and conducted
by local radiology personnel. Customization of CR examination
selections and the appearance of CR images is also a joint
vendor-hospital responsibility.
Figure 2. CR image processed by parameter settings from two different CR scanners in the same facility. This configuration management problem is unique to CR.
|
Fallacy No 3. Use the same radiographic technique for CR that
you used for screen-film.There is no doubt that a screen-film
technique guide can be used to produce CR images. However, it is
unreasonable to expect that a technique guide that was developed
for a specific screen-film system would be optimal for CR,
especially considering the multitude of speed classes for
commercial screen-films. CR can be operated at many different speed
classes. Objective data suggest that in order to match the noise
characteristics of a typical screen-film system, CR must be
operated at approximately 200 speed, corresponding to a medium
speed screen-film. As for kilovolt peak (kVp), CR is reputedly less
sensitive to kVp than most screen-films; however, CR's absorption
"k-edge" is different from most screen-films in commercial use. A
familiar conventional technique guide might be a reasonable place
to begin.
Fallacy No. 4. You cannot cone down with CR. Technologists learn
in their initial training that performing examinations with the
collimator wide open causes more dose to the patient, degrades the
image contrast, and produces more scattered radiation around the
x-ray room. Keeping the collimator wide open is clearly bad
practice in the case of CR, as well. Much of the software
development in CR centered on detection of the collimator
boundaries in images, so that contrast can be maximized for the
region of interest within. Sometimes, especially with un-square,
off-centered collimation or when the projection of one or more
boundaries is interrupted by a solid object such as a pin, or with
multiple fields projected on a single CR plate, the software can
fail to find the boundaries. This can result in bizarre or
nondiagnostic images. These images can usually be recovered by
image processing without needing to repeat the exposure.
Fallacy No. 5. Our CR machines are calibrated at the factory.
Although the CR machines may be calibrated at the end of the
assembly line, much has happened between then and now. CR systems
are complicated electro-optical mechanical systems. There is no
guarantee that vibration and aging during shipment have not
affected alignment and gain of the system. The manufacturer's x-ray
generator may be quite different from that at the hospital. Actual
field experience shows that the performance of CRs varies
substantially from their specifications and that field calibrations
drastically reduce variability between individual machines.
Fallacy No 6. That (exposure indicator) does not mean anything.
One of CR's attributes is the ability to tolerate over- and
under-exposure and still produce an image with consistent
appearance. Unfortunately, CR's adjustable density means that
density is no longer a reliable indicator of exposure factor
control. Each major CR manufacturer (four at this writing) has
developed a numerical indicator of the amount of radiation reaching
the imaging plate. The absolute meaning of the indicator is subject
to interpretation and technical interferences. For example,
interpretation of the indicator depends on proper calibration of
the gain of the CR. These indicators have been demonstrated to
provide reliable measures of exposure to the plate for a given
examination or view, but are more difficult to relate from one type
of examination to another. Software failures, such as
misidentification of collimation boundaries, can also yield
spurious results. Nevertheless, exposure targets and ranges have
been developed using the numerical exposure indicators and, for the
most part, work well for monitoring exposure factor control in
actual clinical practice.
Fallacy No. 7. Use whatever radiographic technique you please
with CR. An even more dangerous fallacy than No. 3, this notion
arises from the observation that the process of photo-stimulable
luminescence on which CR is based produces a signal that is
linearly related to the amount of radiation absorbed by the plate
over at least a factor of 10,000! This ignores other important
facts. First, any electro-optical system, such as CR, has a finite
dynamic range, perhaps 1,000, and a limit of adjustment. Second,
the radiographic projection of real clinical features has a range
of exposures, perhaps a factor of 100 from most to least dense. In
order to capture all the clinical features in a single exposure,
the total tolerance for misexposure is a factor of 10.
Underexposure by a factor of five or likewise overexposure by a
factor of five risks losing clinical features. This description is
not strictly true for CR systems with automatic speed adjustment:
they probably have a tolerance factor of about 50. Not to worry: we
have more than a century of experience controlling the amount of
radiation produced by an x-ray generator, because screen-film is
much less tolerant.
EXPOSURE CREEP CAVEAT
The next important fact is that the amount of noise in the image
depends on the amount of radiation used to produce the image. A CR
system operated at the limit of adjustment produces extremely noisy
images. The more noise in the image, the less image processing that
can be used without producing artifacts. Also, noisier images are
more likely to be rejected by radiologists, who have been trained
to be skeptical of noisy images, which may hide low contrast
details of clinical importance.
On the other hand, using a lot of radiation produces CR images
that are extremely crisp and unlikely to be rejected by a
radiologist. Technologists are quick to notice this, resulting in a
phenomenon known as "exposure creep." Using more radiation than
necessary to produce a diagnostic quality image is clearly a
violation of the "as low as reasonably achievable" (ALARA)
principle and is a disservice to the patient.
Fallacy #8. CR image processing is absolutely optimized for the
visualization of clinical features.
About the only universal agreement that you can find among CR
practitioners is that no one likes the image processing that was
delivered with their particular system. CR image processing was
initially designed to make an unprocessed CR image mimic the
appearance of conventional screen-film when printed by a laser
camera. This approach is reasonable when you consider the worldwide
acceptance of screen-film radiographs for displaying clinical
features. Underlying assumptions are that screen-film rendering is
the best that can be done, that the image is viewed in a static,
not dynamic fashion, and that similar radiographic technique will
be used to produce the images. As CR has been fielded, image
processing has evolved. There is no guarantee that image processing
that was optimal for viewing a transilluminated laser print of a CR
image will also be appropriate for viewing a CR image displayed on
a cathode ray tube (CRT) or flat panel.
Fallacy No. 9. There is no limit to the number of times a CR
imaging plate can be used.
Sure enough, from a physics point of view, there is no limit to
the number of times a CR imaging plate can be exposed to radiation,
have its stored signal harvested, be erased, and exposed again.
Even when bombarded with high energy particles from a cyclotron,
permanent damage to the CR plate is not evident. However, there are
mechanical, chemical, and industrial reasons why plates do not last
forever. The plates are subject to scratches, scuffs, cracks, and
contamination with dust and dirt, which can interfere with the
image. One of the primary components of the plate is iodine, which
oxidizes to iodate on exposure to water. When this occurs, the
plate becomes discolored and the result is artifacts in the image.
Over the years, new formulations have been devised for imaging
plates to enable or exploit the features of new CR scanner
mechanisms. Unfortunately, the modified formulation may be required
by the new mechanism, making the older plates obsolete. Likewise,
the new formulation may not be backward compatible, making the
older mechanism obsolete.
Fallacy No. 10. Any errors in a CR image can be corrected by
image processing.
CR vendors have made dramatic progress in the features and
functions of computer workstations that can be used for quality
control before releasing the images to distribution and archive.
These workstations are especially useful for correcting patient
demographic and examination information as well as verifying proper
exposure indicator values. In contrast to older systems, most
commercially available CRs today retain raw image data to
accommodate reprocessing. However, the discussion of No. 1 above
provides a litany of errors in CR images that cannot be corrected
by image processing. Image processing cannot improve the visibility
of clinical features that are just not present in the raw image,
because they are outside the radiation field or off the edge of the
imaging plate, or because the examination was performed with such
an inappropriate technique that there was no contrast between the
clinical feature and the surrounding tissue. Image processing is a
poor substitute for proper examination technique.
CONCLUSIONS
The important lesson to be learned is that CR is not a panacea.
CR is just another tool for doing the business of radiology. We can
choose to do this business in a professional manner or in a less
efficient, haphazard fashion. Errors can still arise in the
practice of CR, and QC processes must be in place to detect and
correct those errors when they occur. These QC processes are
strikingly similar to the processes we used routinely in
conventional screen-film radiology.
BIBLIOGRAPHY
- Freedman M, Pe E, Mun SK, Lo SCB, Nelson M. The potential for
unnecessary patient exposure from the use of storage phosphor
imaging systems. Proceedings of the International Society for
Optical Engineering. 1993;1897:472-479.
- Gur D, Fuhman CR, Feist JH, Slifko R, Peace B. Natural
migration to a higher dose in CR imaging. Eighth European Congress
of Radiology; September 12-17, 1993; Vienna. Abstract 154.
- Honea R, Blado ME, Ma Y. Is reject analysis necessary after
converting to computed radiography? J Digit Imaging. 2002;15(Suppl
1):41-52.
- Huda W, Arreola M, Jing Z. Computed radiography acceptance
testing. Proceedings of the International Society for Optical
Engineering (SPIE). 1995;2432:512-521.
- Huda W, Slone RM, Belden CJ, Williams JL, Cumming WA, Palmer
CK. Mottle on computed radiographs of the chest in pediatric
patients. Radiology. 1996;199:249-252.
- Oestmann JW, Prokop M, Schaefer CM, Galanski M. Hardware and
software artifacts in storage phosphor radiography. Radiographics.
1991;11:795-805.
- Samei E, Seibert JA, Willis CE, Flynn MJ, Mah E, Junck KL.
Performance evaluation of computed radiography systems. Med Phys.
2001;28:361-371.
- Seibert JA. Photostimulable phosphor system acceptance
testing. In: Seibert JA, Barnes GT, Gould RG, eds. Specification,
Acceptance Testing and Quality Control of Diagnostic X-ray Imaging
Equipment. Woodbury, NY: American Association of Physicists in
Medicine; 1994:771-800. Monograph No. 20.
- Solomon SL, Jost RG, Glazer HS, Sagel SS, Anderson DJ, Molina
PL. Artifacts in computed radiography. AJR Am J Roentgenol.
1991;157:181-185.
- Volpe JP, Storto ML, Andriole KP, Famsu G. Artifacts in chest
radiography with a third generation computed radiography system.
AJR Am J Roentgenol. 1996;166:653-657.
- Willis CE, Weiser JC, Leckie RG, Romlein J, Norton G.
Optimization and quality control of computed radiography.
Proceedings of the International Society for Optical Engineering
(SPIE).1994;2164:178-185.
- Willis CE, Leckie RG, Carter J, Williamson MP, Scotti SD,
Norton G. Objective measures of quality assurance in a computed
radiography-based radiology department. Proceedings of the
International Society for Optical Engineering (SPIE). 1995;
2432:588-599.
- Willis CE, Mercier J, Patel M. Modification of conventional
quality assurance procedures to accommodate computed radiography.
S/CAR '96 Computer Applications To Assist Radiology. Great Falls,
Va: Society for Computer Applications in Radiology;
1996:275-281.
- Willis CE. Computed radiography imaging and artifacts. In:
Siegel EL, Kolodner RM, eds. Filmless Radiology. New York:
Springer-Verlag; 1999:137-154.
- Willis CE. Computed radiography: QA/QC. In: Practical Digital
Imaging and PACS. Medical Physics Monograph No. 28. Madison, Wis:
Medical Physics Publishing; 1999b:157-175.
- Willis CE. Quality control of computed radiography systems.
Quality assurance for the third millennium. American Association of
Physicists in Medicine Southwest Regional Chapter Spring Meeting
and Conference; May 2, 2000; Puerta Vallarta, Mexico.
- Willis CE. Quality assurance: an overview of quality assurance
and quality control in the digital imaging department. In: Quality
Assurance: Meeting the Challenge in the Digital Medical Enterprise.
Great Falls, Va: Society for Computer Applications in Radiology;
2002:1-8.
Charles E. Willis, PhD, is a medical physicist at Texas Childrens Hospital, Houston, and an associate professor at Baylor College of Medicine.