In radiology departments with strong physicist support and a high volume of brain, neurological, and musculoskeletal imaging, 3T MRI will likely add value
A MEDLINE search for "high-field MRI" retrieves mainly studies
performed at 1.5 tesla. That may soon change. In 1998, the US Food
and Drug Administration gave marketing clearance for scanners
operating at up to 4T, and in 2002, the agency approved some 3T
scanners for the brain or the whole body. More than 100 of these
machines are now installed, and they are likely to become the
standard.
But when? Do you need a 3T scanner today? There are clearly
studies that can be performed at 3T that are not possible, or at
least not as easy, at lower field strengths, and many studies can
be accomplished faster. The clinical role of these scanners
vis-ŕ-vis those of lower field strengths, however, is not yet
defined. Although active shielding of the scanners minimizes the
architectural problems, permitting installation in facilities now
equipped with 1.5T scanners, acquisition of a higher-field machine
mandates considerable adjustment of an MRI practice, including
revision of many imaging protocols. Radiologists who use such
scanners also stress the need for close cooperation with a
physicist to get good results.
Is it worth it?
The Promise
The great appeal of 3T MRI is improvement in image quality and
resolution. Because the signal-to-noise ratio (SNR) correlates in
approximately linear fashion with field strength, it is roughly
twice as great at 3T as at 1.5T. The time necessary to acquire
satisfactory images can be substantially reduced, minimizing motion
artifacts and possibly speeding patient turnover. Alternatively,
the same acquisition time can be used to obtain images at higher
resolution. Functional MRI (fMRI) and MR spectroscopy (MRS)
benefit significantly. Also, greater contrast is available at
higher field strength, a fact already well known from comparisons
of images obtained at 0.5T, 1T, and 1.5T.1,2 Among other
benefits, higher contrast may permit reduction of gadolinium doses
and, in some cases, earlier detection of disease, a possible
stimulus for more patient referrals.
But Wait a Minute&
Of course, the greater SNR and contrast come at a price. First,
tissues differ in their magnetic susceptibility, and this effect is
exacerbated at higher field strengths. Indeed, it was long thought
that these differences would preclude clinical use of
ultra-high-field systems. In the brain, the effects are
particularly noticeable on T2-weighted images of iron-containing
tissues and at tissue borders and airbone interfaces such as around
the paranasal sinuses.3 On T1-weighted images,
susceptibility effects can create bright areas that lead to
confusion on postcontrast images if a precontrast view is not
available for comparison. However, this problem is not
insurmountable: the undesirable effects can be reduced by measures
such as increasing the spatial resolution or acquiring thinner
sections.
Second, there are safety concerns. The energy deposited in the
patient's tissues is fourfold higher at 3T than at 1.5T.
Particularly with radiofrequency-intensive pulse sequences such as
fast spin echo and fluid attenuation inversion recovery (FLAIR),
the limit on the specific absorption rate (SAR) power deposition
prescribed by the FDA can easily be reached using protocols that
are satisfactory at 1.5T.3 Scanner manufacturers are
incorporating modified pulse sequences to avoid this problem, which
can also be solved by restricting the volume of tissue that is
studied in detail. Another safety issue is the virtual absence of
data on scanning of patients having various types of implants at
3T.
Third, 3T images are more subject to flow artifacts. Although
these can be reduced by such measures as cardiac gating, there may
remain studies such as routine brain imaging where 1.5T is
superior.
Two other effects of higher field strengths have both drawbacks
and benefits. One is the doubling of the chemical shift when the
field strength is doubled. This effect can cause misregistration of
fat and water, especially after gadolinium administration, leading
to erroneous diagnoses, but it improves the analysis of MR spectra
by increasing the resolution of different metabolites. The second
is the longer and more variable T1 relaxation time. For example, at
3T, relaxation time is as much as 40% greater for brain parenchyma,
whereas there is little change for cerebrospinal fluid.4
A potentially important consequence is a reduction in the contrast
between the white and gray matter in the brain. On the other hand,
the effect can enable better background suppression and reduction
of the in-plane saturation effect for time-of-flight (TOF) magnetic
resonance angiography (MRA).
Finally, there are some practical disadvantages of 3T scanners.
At present, these machines cost at least $1 million more than a
1.5T scanner, although this is likely to change as more 3T systems
are installed. Present magnets also have smaller bores than those
on 1.5T systems, making them too small to image some patients and
increasing the likelihood that an examination will be limited by
patient claustrophobia.
What Is Being Done at 3T?
Success with cochlear implants requires high-resolution imaging
to determine the correct length and dimensions.5 At the
Mayo Clinic in Rochester, Minn, the modality of choice is a
three-dimensional (3D) fast spin echo MRI scan at 3T with a hybrid
array coil that provides as much as a 250% improvement in SNR
compared with a standard head coil.6 In this way, it is
possible to distinguish the acoustic from the vestibular branches
of the VIIIth cranial nerve, and the anatomy of the labyrinth can
be examined with maximum intensity projection reformatting. The
Clinic is correlating audiometric data with the extent of cochlear
nerve atrophy depicted by MRI in the hope of being able to
determine which ear would benefit most from an implant.
Higher-field scanners promise to deliver further improvements in
MRA by creating greater vessel conspicuity and reducing artifacts
by shortening the acquisition time.7 A team at the Mayo
Clinic compared 3D TOF 1.5T and 3T for MRA of the intracranial and
cervical vessels in 20 patients, with two radiologists interpreting
the studies independently.8 Voxels as small as 0.62
mm3 were readily obtained at 3T, and the images were
rated significantly better. In the neck, it was possible to
identify carotid dissections and ulcerations of atherosclerotic
plaque.
Of course, the greatest demand for MRA is in the coronary
system, which has resisted noninvasive angiography because of the
small size and nearly constant motion of the vessels. There is some
evidence that 3T will make such studies possible. In a feasibility
study, a team from Beth Israel Deaconess Medical Center and Harvard
Medical School scanned nine healthy adult volunteers.9
Two scout scans were acquired, and a navigator echo was positioned
at the liverheart boundary to allow subsequent images to be
acquired during free breathing. A 3D segmented k-space gradient
echo sequence was then used to acquire 10 slices with an effective
thickness of 3 mm and an in-plane resolution of 0.7-1.0 mm over an
average time per scan of about 7 minutes and a total session of no
more than 1 hour. A series of 20 1.5-mm slices were then
interpolated, and the 3D images were analyzed. The images were of
high quality in all subjects, and it was possible to see small
branching vessels as well as proximal and mid coronary
arteries.
The musculoskeletal system is one of the most common
applications of MRI, and there is evidence that 3T scanners will
find a place there. At the 2003 Scientific Meeting of the
Radiological Society of North America, a research team from Zurich,
Switzerland, reported on a direct comparison of 1.5T and 3T imaging
for the wrist.10 The higher field strength provided
superior bonemuscle and bonecartilage contrast and better
visibility of the intercarpal ligaments and cartilage and the
median and ulnar nerves. The principal drawback was the much longer
scan time required: an average of 56.2 vs 27.7 minutes.
Numerous applications are being developed for the brain, and
some clinicians believe 3T should always be used for this organ if
available. A German group compared 1.5T and 3T in 11 consecutive
patients with multiple sclerosis, with the latter field strength
being exploited to obtain more and thinner slices in the same scan
time required at 1.5T.11 Images acquired at 3T were
rated significantly superior in lesion conspicuity and diagnostic
value, although artifacts were more common. In these investigators'
view, 3T MRI "may further strengthen the role of MRI as the most
sensitive paraclinical test available for the early diagnosis of
multiple sclerosis." Researchers at the University of Cambridge in
the UK used diffusion tensor imaging (DTI) at 3T in 20 patients
with known primary brain tumors or metastases and calculated
relative anisotropy indexes, a measure of tissue
disorganization.12 In 10 of the 13 patients with
high-grade gliomas, the lesions measured by DTI were larger than
those seen on T2-weighted images, and previously unrecognized
contralateral abnormalities were found by DTI in the white matter
in four of the 13 patients. The team suggested that such
abnormalities indicated infiltration by tumor and that DTI "may
provide a useful method of detecting occult white matter invasion
by gliomas."
The superiority of 3T imaging for MRS has resulted in its use in
numerous clinical and investigational studies of the brain. At
Albert Einstein College of Medicine in New York, proton MRS at 3T
is used to define the site and extent of temporal lobe epilepsy
foci for the surgeons. Investigators at the National Cardiovascular
Center in Osaka, Japan, have applied proton MRS to studies of the
metabolic abnormalities of Alzheimer's disease.13 A
particularly interesting application was described at the RSNA by
Richard Frey, MD, and coworkers from Vienna, Austria.14
In a double-blind study, those investigators used single-voxel
proton MRS at 3T in 22 depressed patients and 22 age- and
sex-matched control subjects, with the patients being imaged before
and after treatment with citalopram, a specific serotonin-reuptake
inhibitor. Concentrations of several compounds, including
creatine, choline, and myoinositol, were measured. Before
treatment, depressed patients had significantly higher
concentrations of creatine in the frontal lobes than did the
control subjects; with drug treatment, their creatine
concentrations became indistinguishable from those of the controls.
The investigators concluded that MRS at 3T can be used to study the
metabolic changes of depression and the effects of treatment.
Conclusion
Although it may be true, as some have asserted, that the
question about installing 3T MRI is not if but when, each
department will have to consider whether it needs the higher-field
scanners enough to make the required adjustments. The new scanners
are expected to have their greatest utility in applications such as
MRS, TOF MRA, fMRI, and DTI. Until the role of 3T for other
purposes is better defined, departments that do not perform these
high-end studies may prefer to wait.
Judith Gunn Bronson, MS, is a contributing writer for Decisions in Imaging Economics.
References:
- Elster AD. How much contrast is enough? Dependence of enhancement on field strength and MR pulse sequence. Eur Radiol. 1997;7(suppl 5):276 - 280.
- Chang KH, Ra DG, Han MH, et al. Contrast enhancement of brain tumors at different MR field strengths: comparison of 0.5T and 2.0T. Am J Neuroradiol. 1994;15:1413 - 1419.
- Thulborn KR, Davis D. Clinical MRI at 3.0 Tesla: performance and safety. Proc Int Soc Magn Reson Med. 1999;7:828.
- Lin C, Bernstein MA, Huston J, Fain S. Measurements of T1 relaxation times at 3T: implications for clinical MRA. Proc Int Soc Magn Reson Med. 2001;9:1391.
- Witte RJ, Lane JI, Driscoll CL, et al. Pediatric and adult cochlear implantation. Radiographics. 2003;23:1185 - 1200.
- Kocharian A, Lane JI, Bernstein MA, et al. Hybrid phase array for improved internal auditory canal imaging at 3.0-T MR. J Magn Reson Imaging. 2002;16:300 - 304.
- Campeau NG, Huston J III, Bernstein MA, Lin C, Gibbs GF. Magnetic resonance angiography at 3.0T: initial clinical experience. Top Magn Reson Imaging. 2001;12:183 - 204.
- Bernstein MA, Huston J III, Lin C, Gibbs GF, Felmlee JP. High-resolution intracranial and cervical MRA at 3.0T: technical considerations and initial experience. Magn Reson Med. 2001;46: 955 - 962.
- Stuber M, Botnar RM, Fischer SE, et al. Preliminary report on in vivo coronary MRI at 3 Tesla in humans. Magn Reson Med. 2002;48:425 - 429.
- Saupe N, Luechinger R, Boesiger P, Marincek B, Weishaupt D. MR imaging of the wrist at 3T: comparison between 1.5T and 3.0T [abstract 1374]. Presented at: Radiological Society of North America RSNA 2003 89th Scientific Assembly and Meeting; December 4, 2003; Chicago.
- Bachmann R, Reilmann R, Kraemer S, Allkemper T, Kugel H, Heindel W. Multiple sclerosis: comparative MR-imaging at 1.5 and 3.0 Tesla [abstract 1465]. Presented at: Radiological Society of North America RSNA 2003 89th Scientific Assembly and Meeting; December 5, 2003; Chicago.
- Price SJ, Burnet NG, Donovan T, et al. Diffusion tensor imaging of brain tumours at 3T: a potential tool for assessing white matter tract invasion? Clin Radiol. 2003;58:455 - 462.
- Hattori N, Abe K, Sakoda S, Sawada T. Proton MR spectroscopic study at 3 Tesla on glutamate/glutamine in Alzheimer's disease. Neuroreport. 2002;13:183-186.
- Gruber S, Frey R, Stadlbauer A, Moser E. Single voxel proton spectroscopy at 3 Tesla in depressive patients before and after treatment: a controlled double-blind design [abstract 802]. Presented at: Radiological Society of North America RSNA 2003 89th Scientific Assembly and Meeting; December 2, 2003; Chicago.